What Is Quality Assurance?






  • AAPM definition of QA: “To assure that machine characteristics do not deviate significantly from their baseline values acquired at the time of testing and commissioning.”



    • Linac measurements are taken during acceptance and commissioning.


    • The AAPM requires daily, monthly, and annual QA tests to ensure that linacs are able to deliver treatment as planned.


    • TG40 set standards for basic linacs without MLCs, on-board imaging, or respiratory gating.


    • TG142 adds standards for MLCs, IMRT, SBRT, on-board imaging and respiratory gating.


  • QA intervals and instrumentation:



    • Daily QA may be done using relatively imprecise “spot check” instruments such as a daily QA device.


    • Monthly QA should use more precise instrumentation than daily QA, as determined by the physicist.


    • Annual QA should use the most precise commissioned instruments, including TG51 calibration with National Institute for Standards and Technology (NIST)-traceable ion chambers.


  • Tolerance level and action level:



    • The tolerance levels are considered “acceptable” for clinical treatment.



      • Tolerances are typically 3 % for daily checks, 2 % for monthly/annual checks, or 1 % for stereotactic.


      • If the equipment consistently violates tolerance levels, treatment should be stopped until the equipment is brought into compliance.


    • However, daily output checks may be more variable due to their rapid and imprecise nature.



      • An action level of 5 % is recommended for stopping treatment after a daily output check deviation.


      • If the deviation is between 3 and 5 %, treatment may continue “for the short term” until the discrepancy can be addressed.