Sonographic Technique and Cross-Correlation with Mammography

2 Sonographic Technique and Cross-Correlation with Mammography


The main applications of ultrasound are identification of a palpable breast lump and clarification of a confusing mammographic finding. In the past, a common reason that sonographic applications were limited was that the ultrasound examiner could not cross-correlate the physical examination or the mammographic findings with the sonographic information. This inability to cross-correlate the ultrasound examination with the physical examination or mammographic asymmetry is frustrating and leads to long, sonographic examinations. To develop the ability to cross-correlate sonography with mammography, you should have optimal sonographic equipment and technique, be able to palpate breast masses, be familiar with normal mammographic and sonographic breast anatomy, recognize mammographic and sonographic signs of malignancy, and apply anatomical knowledge to sonographic scanning.


image Ultrasound Equipment and Technique


Equipment


In the past, breast ultrasound has required the least sophisticated equipment because there were few sonographic breast applications, and these applications required only simple equipment. However, if you wish to have a high rate of localizing solid masses as well as an optimal image to characterize the mass, then you need a sophisticated machine. In this book, when I refer to high frequency, I am generally referring to imaging with frequencies ≥ 10 MHz. To have maximal flexibility, you should have a machine that has linear transducers with frequencies ranging from 7 to 15 MHz.


High frequency is important, as many breast structures are small. An important aspect of the technical advancement of mammography is the improvement of spatial resolution. Because of high mammographic resolution, mammographers are identifying smaller and subtler abnormalities. To be an effective adjunctive test, high sono-graphic spatial resolution is needed to clarify these subtle findings. Because normal breast structures such as ducts and terminal duct lobular units are small, high spatial resolution allows the ultrasound examiner to quickly recognize normal breast architecture and identify small malignant masses in the ductal system.


Besides being able to identify smaller malignancies, high spatial resolution improves the sonographic image so one can better characterize masses. This improved image is similar to the visual effect experienced by a nearsighted person who starts wearing glasses. The image is sharper, and subtle or smaller details are clearer (Fig. 2.1; see also Case 8.8). For breast malignancies, the most important information produced by high-frequency ultrasound is the improved ability to see the margin of the mass and identify secondary signs of malignancy, such as spiculation and architectural distortion. Occasionally, you can trace the path of the malignancy through the ductal system and identify the extent of the disease better than with mammography (see Cases 31.1 and 31.2).


image Imaging Techniques


Besides the availability of the high-frequency transducers, sophisticated ultrasound machines have more technical factors that can be manipulated to improve imaging quality. For breast ultrasound, an important factor in image quality is being able to obtain excellent contrast resolution. The reason contrast resolution is important is that you must be able to distinguish a variety of masses from the background parenchyma. When the breast is fatty, focal masses such as complex cysts, lymph nodes, fibroad-enoma, and cancers may be difficult to identify. When the breast is dense, fat necrosis and surgical or radial scars may be hard to locate. Many factors can improve contrast resolution. High spatial resolution produced by high-frequency imaging improves contrast resolution because the assignment of gray shades is more precise. Reducing the dynamic range may improve contrast resolution, as this method exaggerates differences between the gray shades of structures. A variety of proprietary postprocessing programs improve contrast resolution by enhancement of specified gray shades on a point-by-point basis. One type of proprietary postprocessing program improves contrast resolution by enhancement of specified gray shades on a region-by-region basis. You should consult the manufacturer of the equipment to learn which imaging techniques are available (Fig. 2.2; also see Case 6.11).


image


Fig. 2.1 (A) Right mediolateral oblique (MLO) mammo-gram. (B) Right craniocaudal (CC) mammogram. (A,B) In the upper outer right breast, there is a lobulated density (circle). (C) Right radial breast sonogram: a 5 MHz transducer does not clearly demonstrate the lobulated mass (arrows). The inadequate size and contrast resolution result in poor definition of the mass. (D) Right radial breast sonogram: an 8 MHz transducer improves the definition of the mass. This result allows an observer to confidently localize the mass. This mass is a fibroadenoma.


Besides adjusting image contrast, you should be aware of software methods to optimize resolution. These methods include increasing the line density of the image, increasing the persistence, and adjusting the focal zones. The main disadvantage of these methods is a slower frame rate. If you are merely characterizing a lesion, a slower frame rate may not be a problem. However, the slower frame rate may be disconcerting with real-time imaging of interventional procedures.


Color or power Doppler is a useful method to quickly assess vascularity. Breast vascularity is low, so you should be aware of methods to optimize the color or power Doppler. Generally, this means that you are using a color or power Doppler frequency slightly lower than the gray scale frequency and the focal zone adjusted at the correct depth. The filter and scale should be low. The Doppler gain is optimized by initially increasing the gain until the entire screen is filled with color and then by slowly reducing the gain until the color appears only within pulsating vascular structures. If no color is detected using these methods, then the sample volume size should be increased. Increasing the sample volume size reduces color resolution. The color may “bleed” and be demonstrated outside the vessel walls. Color or power Doppler is useful to delineate vessels or highly vascular structures such as arteriovenous malformations. This Doppler technique is also useful to clarify whether a hypoechoic or anechoic mass is cystic or solid (Fig. 2.3; also see Case 6.25).


image


Fig. 2.2 (A) Left MLO mammogram. (B) Left CC mammogram. (C) Left MLO spot magnification mammogram. (A–C) In the left upper outer quadrant, there is a small, irregular mass (circle). (D) Left antira-dial (8 MHz) breast sonogram: sonographic examination of the mammographic mass with a low-contrast technique poorly demonstrates the mass (arrows). The hypoechogenicity of the mass blends into the surrounding fat. (E) Left antiradial (8 MHz) breast sonogram: sono-graphic examination of the same location as (D) with high-contrast technique greatly improves the conspicuity of the mass. This mass is a mucinous carcinoma.


Dynamic clips are useful to document vascularity and to demonstrate the spatial relationship of multiple lesions. Dynamic clips are the ideal method to show color flow in pseudoaneurysms or intravenous contrast enhancement of solid masses. Until high-resolution three-dimensional (3D) imaging is universally available, dynamic clips are an excellent way to demonstrate the relationship of multiple cysts to a solid mass or to show debris or calcifications moving within a complex cyst.


Wide field-of-view compound imaging is sometimes useful to document larger masses or the relationship of multiple masses in the same plane. The wider field of view provides observers with more landmarks, so cross-correlation with mammography and magnetic resonance imaging (MRI) may be easier.


Newer sonographic techniques that may have more applications in the future are 3D imaging and harmonic imaging. Like compound imaging, 3D imaging may produce a wide field of view that would be similar to a mammogram or MRI. In the future, 3D imaging may also allow surgeons and patients to better appreciate the location and size of sonographic findings and facilitate surgical planning.


Harmonic imaging may improve image resolution and increase both gray scale and color Doppler sensitivity for intravenous sonographic contrast agents. These contrast agents may improve both vascular and gray scale characterization of masses.


image Approach to a Palpable Mass


In many breast centers, palpable masses are the most common reason for a breast sonogram. Therefore, it is important that sonographic breast imagers learn how to palpate breasts. Usually, the patient will be able to identify the palpable lump. When the patient locates the lump, the breast imager should confirm the presence of the lump by palpating the area identified by the patient. Even if the lump is obvious, the imager should scan the lump and reconfirm the location of the lump by moving a finger into the scan plane. If the imager cannot detect the lump, or if the patient is not sure of the exact location of the lump, then palpation of the entire quadrant is useful. By palpating a larger area, the imager is able to detect asymmetries within the region. Finally, if palpating the quadrant is not helpful, palpating the comparable area in the opposite breast is helpful. Commonly, the parenchymal pattern of patients is symmetric, so the physical exam is also symmetric. By palpating the corresponding contralateral quadrant, you can detect abnormal asymmetries. This technique is particularly useful with malignancies that are commonly difficult to feel, such as lobular carcinoma.


image


Fig. 2.3 (A) Right MLO mammogram. (B) Right CC mammogram. (A,B) The patient and the breast surgeon have identified some small, palpable lumps at the 9 o’clock position of the right breast. The lumps are arranged in a linear pattern. The left breast has similar lumps that are less conspicuous. Bilateral mammograms are normal. (C) Right radial breast sonogram: gray scale sonographic examination of the palpable lumps shows normal tissue. (D) Right radial breast sonogram: color Doppler examination of the palpable lumps demonstrates that the lumps are due to the lateral blood vessels of the breast. Each lump corresponds to the superficial curve of the blood vessel (arrows).


image Cross-Correlation of Sonographic and Mammographic Image


To accurately, effciently, and confidently identify a mammographic abnormality sonographically, the technician who performs the breast sonographic examination should be familiar with mammographic imaging. Furthermore, the ultrasound examiner should be able to review the mammogram and identify internal landmarks that can be cross-correlated with the ultrasound. Finally, by confirming the mammographic landmarks sonographically, the examiner should be able to pinpoint the location of the mammographic abnormality in the breast with ultrasound and consequently be able to explain the etiology of the puzzling mammographic finding.


Unfortunately, sometimes the ultrasound examiner does not attempt to closely cross-correlate anatomically the ultrasound examination with the mammogram. Some examiners do not attempt to correlate the exams because they do not routinely interpret mammograms and are uncomfortable reviewing mammograms. However, more common reasons for lack of close cross-correlation include the following: (1) The sonographic image has a small field of view compared with the mammographic global field of view. (2) The patient position for an ultrasound examination is completely different from the position for the mammographic examination. Therefore, the position of a breast mass for these exams appears extremely different. (3) Even if the examiner places the patient in the same position, the difference in technology between ultrasound and mammography creates different orientations of tissue visualization. (4) Unlike other organs, the breast does not have a uniform or constant normal anatomy. The breasts of different individuals have different breast architecture. Furthermore, some individuals have a right breast that exhibits a pattern different from the left breast. Finally, the breasts of many individuals change with age.