Common Errors in the Doppler Ultrasound Display of Uterine Blood Flow and Fetal Vessels






6


Common Errors in the Doppler Ultrasound Display of Uterine Blood Flow and Fetal Vessels


Correct Display of Vessels with Normal Instrument Settings


image

Fig. 6.1 Uterine a. Color Doppler display of the uterine a. crossing over the iliac vessels. Acute angle of incidence, indication of a notch in the spectrum.


image

Fig. 6.2 Optimal method of examining the umbilical a. The umbilical a. is shown with forward flow, the umbilical v. with reverse flow (upper margin of display).


image

Fig. 6.3 Fetal aorta. Acute angle, legend at the margin of the display, wall filter and pulse repetition frequency (PRF) at optimal setting. Spectrum unremarkable.


image

Fig. 6.4 Middle cerebral a. Angle almost 0°, optimal PRF, and sample time duration optimizes the flow profile. Unremarkable spectrum.


Role of the Angle in the Doppler Examination


image

Fig. 6.5 Uterine a. Displayed is the origin of the vessel on the other side of the iliac crossing. The Doppler angle is more obtuse than in Fig 6.1. The resistance index (RI) is elevated and there is a visible notch.


image

Fig. 6.6a Display of the umbilical a. Because the vessel is poorly focused, the spectrum is not homogenous, the RI is elevated, and the umbilical v., shown in reverse flow, is displayed only partially.


image

Fig. 6.6b The umbilical aa. are displayed three times in this spectrum: 1. Reverse flow, 2. Forward flow with a high RI, 3. Forward flow with a low RI. When a choice must be made between several such curves, the problems posed by the examination of umbilical aa. become clear.


image

Fig. 6.7a The fetal aorta has been examined with a poor angle of almost 90° in the center of the display. The RI is almost 85%.


image

Fig. 6.7b The same aorta as in Fig. 6.7a examined a little later with an acute angle and at the margin of the display. Result: Normal spectrum, unremarkable RI of ca. 75%.


image

Fig. 6.8a Display of a fetal aorta examined with a 90° angle. The incorrect examination technique results in zero flow in the aorta.


image

Fig. 6.8b The same vessel examined a few seconds later with a Doppler angle of 45°. Result: Normal flow profile; normal RI.


Possible Sources of Error in Doppler Ultrasound Examinations of Maternal and Fetal Vessels


Displaying the Uterine Artery


image

Fig. 6.9a Typical flow profile of a uterine a. with an elevated diastole and a RI of about 50%.


image

Fig. 6.9b Same vessels as in Fig. 6.9a. Error: PRF is set too low, with the result that no Doppler spectrum is displayed.


image

Fig. 6.9c The PRF is adapted, the Doppler spectrum appears. However, systole continues to be incomplete and appears as a “peak” above the baseline.


image

Fig. 6.9d In this case the spectrum of the uterine vessel is displayed optimally; the waveform can be evaluated.


image

Fig. 6.9e A venous uterine vessel appears in reverse flow.


image

Fig. 6.10a Uterine a. The vessel has been detected in B-mode by color Doppler and has been captured by the Doppler beam at an acute angle. The right-hand spectrum cannot be evaluated: The baseline is too elevated and the waveform has been cut off by the upper margin of the display. The waveform displayed by the instrument suggests an elevated systole, but this cannot be confirmed visually. In this case the waveform is somewhat unfavorable, not very helpful.


image

Fig. 6.10b Uterine a. Error: Baseline too high.


image

Fig. 6.10c Error in the display of the uterine a. The waveform displays faulty curves, due to too low amplification. Systole is incomplete.


image

Fig. 6.10d The spectrum displaying the flow of the uterine a. is complete here. The waveform represents the course of diastole correctly.


image

Fig. 6.11a Notched uterine a.


image

Fig. 6.11b Incomplete display ofthe systole of a uterine a. PRF set too low. Notch.


image

Fig. 6.11c Same vessel as in Fig. 6.11 b. PRF was raised: Systole is now displayed completely.


Displaying the Umbilical Artery


image

Fig. 6.12a Display of the umbilical a. Error: The waveform cannot be interpreted, since neither systole nor diastole was recorded uniformly. The Doppler beam did not encompass the coil of the umbilical cord in continuity.


image

Fig. 6.12b The umbilical cord has been encompassed in continuity, the spectrum is continuous. Error: The wall filter is too high. Systole is cut off. Possible corrective actions: Lower baseline; reduce wall filter.


image

Fig. 6.13 a, b Display of umbilical a. Each figure shows two coils of umbilical cord caught by the Doppler beam. As a result there is a tracing on the baseline with forward flow, and a second curve at the upper margin of the display with reverse flow. When the waveform is not homogeneous, it is not possible to derive Doppler parameters.


image

Fig. 6.14 Umbilical a. with characteristic sawtooth pattern. In the first part of the tracing the spectrum is inverted.


image

Fig. 6.15a Umbilical a. The spectrum is inverted, systolic peaks are cut off and appear at the upper margin of the display.


image

Fig. 6.15b The waveforms in Fig. 6.15a have been corrected. The angle is set too high, resulting in a curve that does not touch the baseline.


image

Fig. 6. 16 Umbilical a. Background noise is too strong in this examination of the vessel. Sample duration too high.


Displaying the Fetal Aorta


image

Fig. 6.17 Display of fetal aorta. Error: The B-mode display is indistinct. The curve is incomplete. The faulty choice of a too high wall filter leads to an erroneous impression of zero flow.


image

Fig. 6.18 Display of fetal aorta. Error: Angle ca. 80°. Result: The RI of the curve displayed will be abnormal by being falsely elevated, since the Doppler angle is not <60°.


image

Fig. 6.19 Fetal aorta. The vessel has been examined in the center of the display, rendering the recording of the spectrum unproductive.


image

Fig. 6.20 Fetal aorta displayed with an obtuse Doppler angle. RI: 76%, unremarkable. The correct examination appears at the left margin of the display.


image

Fig. 6.21a Fetal aorta. The sample duration has been set too high, with the result that the diastole cannot be viewed and noted. The tracing does not lie on the baseline, indicating faulty filter setting.



image

Fig. 6.21c In this display the baseline is set too high and has been corrected upward in error. The result is that systole has been cutoff and the peaks appear at the bottom of the display.


image

Fig. 6.21d The baseline has been corrected to its proper level. The filter is still too high; the B-mode image blurry.


image

Fig. 6.22 Fetal aorta. Error: Strong background noise blurs the contour of the Doppler spectrum. Correction: Reduced amplification.


image

Fig. 6.23 Fetal aorta. Error: The waveform tracing shows reverse flow. Systole lacks homogeneity. A venous vessel is seen in forward flow. B-mode image shows a poorly selected Doppler angle. Correction: Invert the angle.


image

Fig. 6.24 Fetal aorta. Error: Tracing shows reverse flow. Faulty angle, resulting in falsely showing reverse flow in the vessel. Diastole can only be presumed at the upper margin of the display. Correction: Inversion, adaptation of the PRF, correction of the angle of incidence.


image

Fig. 6.25 Doppler signal in the fetal aorta during arrhythmia. Doppler parameters in this situation are inadmissible. Conclusions are only valid for orientation.


image

Fig. 6.26 Fetal aorta. Error: Angle reading in the B-mode image is incorrect. 60° was set, but this does not follow the actual course of the aorta. The spectrum is inverted; baseline shift has not been utilized.


image

Fig. 6.27 Fetal aorta: Error: Sample duration set too high. Systole cannot be delimited. Angle not set optimally. Correction: Reduce sample rate, if necessary raise PRF.


Displaying the Middle Cerebral Artery


image

Fig. 6.28 Middle cerebral a. with diastolic flow. The result is a low RI of 74%. The angle is correctly acute.


image

Fig. 6.29a Middle cerebral a. Correct display: Optimal angle, PRF, and sample duration set correctly. Systole and diastole well displayed. RI unremarkable.


image

Fig. 6.29b Incorrect display. Spectrum inverted, baseline shifted in wrong direction, systolic peaks appear at the upper margin of the display.


image

Fig. 6.29c Baseline and direction of flow incorrect. Correction: Shift baseline downward and invert.

Stay updated, free articles. Join our Telegram channel

Jan 10, 2016 | Posted by in OBSTETRICS & GYNAECOLOGY IMAGING | Comments Off on Common Errors in the Doppler Ultrasound Display of Uterine Blood Flow and Fetal Vessels

Full access? Get Clinical Tree

Get Clinical Tree app for offline access