Exposure and timing circuits

Chapter 29 Exposure and timing circuits





29.2 Preparation for exposure


Before the X-ray tube can deliver an exposure for a predetermined time interval, it must first be prepared for exposure. This is done during the ‘prep’ stage of the exposure sequence and during this time two major things happen:



If an exposure is made before these processes are completed, there is a risk that the incorrect mA will be delivered or that the target of the anode may be subjected to localized overheating, resulting in damage to the anode. In Sections 13.5 and 13.6, it was shown that selection of a suitable resistance would determine the time taken to charge or discharge a capacitor. This principle is used to provide a delay function prior to the X-ray exposure.


Initiating the prepare sequence causes a switch to open, ‘shorting’ the capacitor to remove any residual charge present. Current flows in the stator circuits and the anode commences to rotate. The switch then closes, which permits a direct current (DC) supply to charge the capacitor through the resistor. When the capacitor is charged, it operates an electronic switch in the exposure circuits, permitting the exposure to commence.


There are two circuit sections responsible for the actual exposure:




29.3 The switching section


The function of the switching section is to connect the high voltage (kVp) to the X-ray tube during the exposure and to disconnect this supply from the tube at the end of the exposure. Such switching commonly occurs between the autotransformer and the high-tension transformer, where it is known as primary switching, or between the high-tension transformer and the X-ray tube, where it is known as secondary switching.



29.3.1 Primary switching


All modem X-ray units make use of solid-state switching. This type of switching has the advantage that there are no moving parts, overcoming the problems experienced with earlier mechanical systems. A simple circuit containing a solid-state switching system is shown in Figure 29.1. Silicon-controlled rectifiers (SCRs), a type of thyristor, are used for this purpose. Two thyristors (connected in inverse parallel) are required to switch an alternating current (AC) as each conducts the half-cycle when that SCR is forward biased. At the end of each half-cycle, each SCR will cease to conduct as the potential difference across it drops to zero and so a voltage pulse must be applied to its gate if it is required to conduct during the next half-cycle.



As an alternative to the SCR, a triac may be used. This device acts as two SCRs connected in inverse parallel and, if pulsed with an alternating supply, will conduct in both phases of the AC cycle. Like the SCR, the device will only conduct when the voltage is not at zero volts and the device has been pulsed.


During the exposure, the timer is simply required to apply a sequence of synchronized pulses to the gate of the device at a time slightly later than the mains zero to switch them back on and ensure their continued conduction. At the end of the exposure, these pulses stop and conduction through the device stops at the end of the next half-cycle. The system allows accuracy of one voltage pulse (i.e. an exposure time of 0.01 second in the case of a two-pulse unit, or 0.002 seconds in the case of a medium-frequency unit; see Ch. 28).


Mar 6, 2016 | Posted by in GENERAL RADIOLOGY | Comments Off on Exposure and timing circuits

Full access? Get Clinical Tree

Get Clinical Tree app for offline access