Introduction to skeletal, chest and abdominal radiography

Chapter 4 Introduction to skeletal, chest and abdominal radiography



To avoid repetition within the text, some safety, terminological and technical issues can be addressed by the use of initial statements regarding projection names, patient preparation, selection and use of image recording media, dose reduction methods, image identification and anatomical markers. This information is covered in this chapter.






Image recording (CR cassettes and digital plates)


With the current situation of image recording, where there is a choice of computed radiography (CR), digital radiography (DR) and film/screen systems for use, it has been difficult to select a method of description that accurately embraces the use of all of these methods. It must be said that, in the UK, use of film/screen radiography is now almost obsolete. As CR uses cassettes similar in appearance to conventional film/screen systems, there would appear to be little change from film/screen methods for image receptor (IR) use; however, some notable differences regarding DR exist. It has therefore been decided that the term ‘image receptor’ will be used as an umbrella term. This is intended to include any of the recording systems that may be used by the radiographer. It should also be noted that use of lead rubber for masking is not advisable for CR and DR systems, although some of the positioning images do demonstrate this on a film/screen cassette.


Since the last edition of this book was published there have been significant developments in DR plates and increased use of these. Wireless plates are now widely available, which further improves the flexibility and range of uses for digital plates, with some manufacturers developing DR support units which are flexible in their positioning rather than being fixed vertically or horizontally. DR plates vary slightly in size and are usually square in shape, but generally do not come in the wider range of sizes found with film/screen or CR plates. They can be fixed under a stand or table surface, independent (wired or wireless), in a tray used under the table-top or pulled from the side of the table-top for extremity work.


As a result of the range of possibilities for receptor arrangement, the IR position will simply be referred to as horizontal or vertical and no IR sizes will be given.


DR plates do not require the centre of the body part to be placed coincident with the middle, unlike CR cassette radiography. For this reason, the positioning descriptors provided in this book assume that the radiographer will always ensure that the body part lies within the IR, or within an unexposed section if the IR is used for more than one projection. At times it will be necessary to centre the body part to the middle of the DR plate, e.g. when that body part is large (as in chest or abdomen radiography), and this will be advised in descriptors for some sections, in order to ensure that the whole of the body part is included in the image.


Another point to raise is the use of an antiscatter device (grid), which should be used in conjunction with the IR if scatter reduction is relevant. Their use will be indicated in descriptors when necessary.



Change in terminology for focus film and object film distances


With the disappearance of film/screen radiography it has become necessary to reconsider these radiographic terms in order to ensure accuracy of reference. It has been noted that different terminologies have been introduced in recent years in an attempt to address this issue, and US texts initiated the use of the terms ‘source image distance’ and ‘object image distance’ as long ago as 20052 in an attempt to use more appropriate terms that did not include the word ‘film’. However, one must question the use of the word ‘source’: it is true that the tube target is a source of radiation but use of the word ‘source’ in a radiation environment does imply ‘radioactive source’, simply because ‘source’ is used more routinely when referring to radioactive materials (although it is not inaccurate to refer to electrically produced X radiation as a source of radiation). In addition, use of the word ‘image’ can be considered inaccurate, as the image is latent until digitally processed and displayed. As a result it has been decided to use the terms focus receptor distance (FRD) and object receptor distance (ORD) in this edition; we feel that these are more appropriate, especially as the terms only include one changed word from old terminology, making them more easy to adopt.





Mar 3, 2016 | Posted by in GENERAL RADIOLOGY | Comments Off on Introduction to skeletal, chest and abdominal radiography

Full access? Get Clinical Tree

Get Clinical Tree app for offline access