Physics and Instrumentation in Doppler and B-mode Ultrasonography

2 Physics and Instrumentation in Doppler and B-mode Ultrasonography



This chapter presents an overview of the physical and technical aspects of vascular sonography, including the following: (1) a brief review of relevant ultrasound–soft-tissue interactions, (2) pulse-echo principles and display techniques, (3) harmonic and chirp imaging, (4) the Doppler effect as it applies to vascular sonography, (5) continuous-wave (CW) and pulsed Doppler instrumentation, (6) the common techniques used for displaying Doppler signal spectral information, and (7) extended field-of-view and three-dimensional (3D) techniques.



Sound Propagation in Tissue


Sound waves are produced by vibrating sources, which cause particles in the medium to oscillate, setting up the wave. As sound energy propagates, it is attenuated, scattered, and reflected, producing echoes from various interfaces. In medical ultrasonography, piezoelectric elements inside an ultrasound transducer serve as the source and detector of sound waves. The design of the transducer is such that the waves travel in a beam with a well-defined direction. The reception of reflected and scattered echo signals by the transducer makes possible the production of ultrasound images and allows detection of motion using the Doppler effect. This section inspects factors that are important in the transmission and reflection of ultrasound in tissue.



Speed of Sound


Most ultrasound applications involve transmitting short bursts, or pulses, of sound into the body and receiving echoes from tissue interfaces. The time between transmitting a pulse and receiving an echo is used to determine the depth of the interface. The speed of sound in tissue must be known to apply pulse-echo methods.


Sound propagation speeds depend on the properties of the transmitting medium and not significantly on frequency or wave amplitude. As a general rule, gases, including air, exhibit the lowest sound speed; liquids have an intermediate speed; and firm solids such as glass have very high speeds of sound. Speeds of sound in common media and tissues are listed in Table 2-1. For soft tissues, the average speed of sound has been found to be 1540 m/sec.1 Most diagnostic ultrasound instruments are calibrated with the assumption that the sound beam propagates at this average speed. Slight variations exist in the speed of sound from one tissue to another, but as Table 2-1 indicates, speeds of sound in specific soft tissues deviate only slightly from the assumed average. Adipose tissues have sound speeds that are lower than the average, whereas muscle tissue exhibits a speed of sound that is slightly greater than 1540 m/sec.


TABLE 2-1 Speed of Sound for Biologic Tissue































Tissue Speed of Sound (m/sec) Change from 1540 m/sec (%)
Fat 1450 −5.8
Vitreous humor 1520 −1.3
Liver 1550 +0.6
Blood 1570 +1.9
Muscle 1580 +2.6
Lens of eye 1620 +5.2

From Wells PNT: Propagation of ultrasonic waves through tissues. In Fullerton G, Zagzebski J, editors: Medical physics of CT and ultrasound, New York, 1980, American Institute of Physics, p 381.



Frequency and Wavelength


The number of oscillations per second of the piezoelectric element in the transducer establishes the frequency of the ultrasound wave. Frequency is expressed in cycles per second, or hertz (Hz). Audible sounds are in the range of 30 Hz to 20 kHz. Ultrasound refers to any sound whose frequency is above the audible range (i.e., above 20 kHz). Diagnostic ultrasound applications use frequencies in the 1-MHz to 30-MHz (1 million to 30 million Hz) frequency range. Manufacturers of ultrasound equipment and clinical users strive to use as high a frequency as practical that still allows adequate visualization depth into tissue (see section on attenuation). Higher frequencies are associated with improved spatial detail, or better resolution.


Figure 2-1 shows what might be called a snapshot of a sound wave, captured at an instant of time. It illustrates accompanying compressions and rarefactions in the medium that result from the particle oscillations. The wavelength λ is the distance over which a property of a wave repeats itself. It is defined by the equation




image    (2-1)



where c is the speed of sound and f is the frequency. Table 2-2 presents values for the wavelength in soft tissue, where the speed of sound is taken to be 1540 m/sec, for several frequencies. A good rule of thumb for tissues is the wavelength λt = 1.5 mm/F, where F is the frequency expressed in MHz. For example, if the frequency is 5 MHz, the wavelength in soft tissue is approximately 0.3 mm. Higher frequencies have shorter wavelengths and vice versa.


TABLE 2-2 Wavelengths for Various Ultrasound Frequencies





















Frequency (MHz) Wavelength* (mm)
1 1.54
2.25 0.68
5 0.31
10 0.15
15 0.103

* Assuming a speed of sound of 1540 m/sec.


Wavelength has relevance when describing dimensions of objects, such as reflectors and scatterers in the body. The size of an object is most meaningfully expressed if given relative to the ultrasonic wavelength for the frequency of the sound beam. Similarly, the width of the ultrasound beam from a transducer depends in part on the wavelength. Higher-frequency beams have shorter wavelengths and are narrower than lower-frequency beams.



Amplitude, Intensity, and Power


A sound wave is accompanied by pressure fluctuations in the medium. The pressure profile that could occur for the wave in Figure 2-1 might appear as in the graph in the lower part of this figure. The pressure amplitude is the maximal increase (or decrease) in the pressure caused by the sound wave. The unit for pressure is the pascal (Pa). Pulsed ultrasound scanners can produce peak pressure amplitudes of several million pascals in water when power controls on the machine are adjusted for maximal levels. As a benchmark for comparison, atmospheric pressure is approximately 0.1 MPa, so it is clear that ultrasound fields from medical devices significantly exceed this mark. The high-pressure amplitudes of an ultrasound pulse can easily burst contrast agent bubbles (see later) that are sometimes injected into the bloodstream to enhance echo signals. Diagnostic levels, however, are not believed to create biologic effects in tissues if such gas bodies are not present.


The intensity (I) of a sound wave at a point in the medium is estimated by squaring the pressure amplitude (P) and using I = P2/2ρc, where ρ is the density and c is the speed of sound. Units for ultrasound intensity are watts per meter squared (W/m2) or multiples thereof, such as mW/cm2. In water, a 2-MPa amplitude during the pulse corresponds to a pulse average intensity of 133 W/cm2! This is a high intensity, but, fortunately, it is not sustained by a diagnostic ultrasound device because the duty factor (i.e., the fraction of time the transducer actually emits ultrasound) typically is less than 0.005. Therefore, the time-averaged acoustic intensity from an ultrasound machine, found by averaging over a time that includes transmit pulses as well as the time between pulses, is much lower than the intensity during the pulse. Typical time-averaged intensities at the location in the ultrasound beam where the maximal values are found are on the order of 10 to 20 mW/cm2 for B-mode imaging. Doppler and color flow imaging modes have higher duty factors. Moreover, these modes tend to concentrate the acoustic energy into smaller areas. Time-averaged intensities for Doppler modes may be a few hundred mW/cm2 for color flow imaging and as high as 1000 to 2000 mW/cm2 for pulsed Doppler!2,3


The acoustic power produced by a scanner is the rate at which energy is emitted by the transducer. Average acoustic power levels in diagnostic ultrasonography are low because of the small duty factors used in most equipment. Typical power levels are on the order of 10 to 20 mW for black-and-white imaging, but may be three to four times this value for color flow modes of operation.



Acoustic Output Labels on Machines


The transmit level, or the output power, on most scanners may be adjusted by the operator. Increasing the power applies a more energetic signal to the transducer, thereby increasing the pressure amplitude and increasing the power and the intensity of the waves produced. Higher power levels are advantageous because they enable detection of echoes from more weakly reflecting interfaces in the body. The disadvantage of high power levels is that they expose the tissue to greater amounts of acoustic energy, increasing the potential for biologic effects. Although there are no confirmed effects of ultrasound on patients during diagnostic ultrasound exposures, most operators attempt to follow the ALARA (As Low As Reasonably Achievable) principle when adjusting the power level and other instrument controls that affect output levels.


It would be difficult to follow ALARA without labels on the machine to inform the operator “how much” ultrasound energy is being applied. Although some ultrasound machines display relative output indications, such as a transmit level percentage, a relative level in decibels, or simply the setting of a power control knob, such labels do not provide users sufficient information to help them understand the likelihood that the sound levels produced might be in an undesirable zone.


To help operators implement the ALARA principle, output labels are used that are related to the biologic effects of ultrasound.4 One of the potential effects is “cavitation,” which describes activity of small gas bodies under the action of an ultrasound field. When gas bodies are present, such as when there are contrast agents in the ultrasound field, cavitation increases the local stresses on tissue that are associated with the ultrasound waves. If the wave amplitude is high enough, collapse of the gas body occurs, and this is accompanied by localized energy depositions that significantly exceed depositions that might occur without cavitation. Cavitation is believed to be most closely associated with the peak negative pressure in the ultrasound wave. Scientists have developed a “mechanical index” (MI) that is derived from the peak negative pressure in the medium. For most ultrasound machines, the current maximum MI in the field is displayed in a prominent position on the display (Figure 2-2).



Another way that ultrasound energy may affect tissue is by heating through absorption of the waves. Absorption is one of the mechanisms that result in attenuation of a sound beam as it propagates through tissue. A corresponding index, the “thermal index” (TI) is displayed to indicate the likelihood of heating (see Figure 2-2). This is estimated using the time-averaged acoustic power or the time-averaged intensity, along with detailed mathematical models for the sound beam pattern and assumptions on the ultrasonic and thermal properties of the tissue. Depending on the application, a machine will exhibit either a soft tissue thermal index value (TIs) or a thermal index for the case in which absorbing bone is at the beam focus (TIb). TIc is a thermal index that is used for Transcranial Doppler studies because heating is likely to occur in the cranial bones.


The acoustic output labeling standard calls for a clear display of MI and TI.4 The standard is followed by most ultrasound equipment manufacturers, and it provides ultrasound system operators values of acoustic output quantities that are relevant to the possibility of biologic effects from the ultrasound exposures.



Decibel Notation


Decibels are used frequently to indicate relative power, intensity, and amplitude levels. Their use is a way to express the ratio of two signal amplitudes or two intensities. Suppose one wishes to express how much greater (or smaller) one intensity (I1) is relative to another (I2). Their relative value in decibels is given by



image    (2-2)



Thus, the decibel relation between two intensities is just the log of their ratio multiplied by 10. The same equation holds for expressing the ratio of two power levels. The difference in decibels between two powers is found by taking the log of their ratio and multiplying by 10.


Sometimes amplitudes rather than the intensities of two signals are used to express decibels. For a given decibel level, one must account for the fact that the intensity is proportional to the square of the amplitude. Substituting the corresponding amplitudes (A1 and A2) into Equation 2-2, squaring them, and taking into account that log (x2) is 2(log x), we have the relationship



image    (2-3)



Notice, the multiplicative factor is 20 rather than 10 when converting amplitude ratios to decibels.


Table 2-3 lists decibel values for various intensity and amplitude ratios. Notice that a 3-dB increase in the intensity is the same as doubling the quantity. A 10-dB increase corresponds to a 10-fold increase, and a 20-dB increase means that the intensity is multiplied by 100. The lower half of the table shows decibel changes corresponding to reductions of the intensity. A 3-dB decrease is the same as halving the intensity, and so forth.


TABLE 2-3 Decibel Differences Corresponding to Various Intensity and Amplitude Ratios*



















































Amplitude Ratio (A1/A2) Intensity Ratio (I1/I2) Decibel Difference (dB)
1 1 0
1.41 2 +3
2 4 +6
2.828 8 +9
3.16 10 +10
4.47 20 +13
10 100 +20
100 10,000 +40
1 1 0
0.707 0.5 −3
0.5 0.25 −6

* For example, if I1 is 10 times I2, it is 10 dB greater than I2. A 20-dB difference between two signals corresponds to both a ratio of 10 for their amplitude or a ratio of 100 for their intensities, and so forth.


Frequently, decibels are used to describe the loudness of audible sounds. Here, the level of one sound often is expressed with no explicit comparison to another, such as “the sound intensity of the jet at takeoff was 110 dB.” However, with airborne sounds, a reference intensity is implied when not stated explicitly. This reference is I2 = 10−12 W/m2, the accepted threshold for human hearing.



Attenuation


As a sound beam propagates through tissue, its intensity decreases with increasing distance. This decrease with path length is called attenuation. Attenuation of medical ultrasound beams is caused by reflection and scatter of the waves at boundaries between media having different densities or speeds of sound and absorption of ultrasonic energy by tissues. As mentioned previously, absorption may lead to heating if beam power levels are sufficiently high.


The rate of attenuation in relation to distance is called the attenuation coefficient, expressed in decibels per centimeter. The attenuation coefficient depends on both the medium and the ultrasound frequency. Figure 2-3 illustrates attenuation coefficients for a few tissues, plotted versus the frequency. Attenuation is quite high for muscle and skin, has an intermediate value for large organs such as the liver, and is very low for fluid-filled structures. For the liver, it is approximately 0.5 dB/cm at 1 MHz, whereas for blood, it is about 0.17 dB/cm at 1 MHz. An important characteristic of attenuation is its frequency dependence. For most soft tissues, the attenuation coefficient is nearly proportional to the frequency.1 The attenuation expressed in decibels would roughly double if the frequency were doubled. Thus, higher-frequency sound waves are more severely attenuated than lower-frequency waves, and the high-frequency beams cannot penetrate as far as low-frequency beams. Diagnostic studies with higher-frequency sound beams (7 MHz and above) are usually limited to superficial regions of the body. Lower frequencies (5 MHz and below) must be used for imaging large organs, such as the liver.




Reflection


Figure 2-4 shows an ultrasound image of the carotid artery of a normal adult. The walls of the vessel can be seen because of reflection of sound waves. Echoes from muscle and other tissues are also produced by reflections and by ultrasonic scatter. Both reflection and scatter contribute to the detail seen on clinical ultrasound scans.



Partial reflection of ultrasound waves occurs when they are incident on interfaces separating tissues having different acoustic properties. The fraction of the incident energy that is reflected depends on the acoustic impedances of the tissues forming the interface. The acoustic impedance (Z) is the speed of sound (c) multiplied by the density (ρ) of a tissue. The amplitude or strength of the reflected wave is proportional to the difference between the acoustic impedances of tissues forming the interface.


The reflection coefficient quantifies the relative amplitude of a wave reflected at an interface. It is the ratio of the reflected amplitude to the incident amplitude. For perpendicular incidence of the ultrasound beam on a large, flat interface (Figure 2-5), reflection coefficient (R) is given by




image    (2-4)



where the impedances Z1 and Z2 are identified in Figure 2-5.


Equation 2-4 shows that the larger the difference between impedances Z2 and Z1, the greater will be the amplitude of the echo from an interface, and hence, the less will be the transmitted signal. Large impedance differences are found at tissue-to-air and tissue-to-bone interfaces. In fact, such interfaces are nearly impenetrable to an ultrasound beam. In contrast, significantly weaker echoes originate at interfaces formed by two soft tissues because, generally, there is not a large difference in impedance between soft tissues.5


Large, smooth interfaces, such as those indicated in Figure 2-5, are called specular reflectors. The direction in which the reflected wave travels after striking a specular reflector is highly dependent on the orientation of the interface with respect to the sound beam. The wave is reflected back toward the source only when the incident beam is perpendicular or nearly perpendicular to the reflector. The amplitude of an echo detected from a specular reflector thus also depends on the orientation of the reflector with respect to the sound beam direction. The ultrasound image in Figure 2-4 was obtained using a linear array probe, which sends individual ultrasound beams into the scanned region in a vertical direction as viewed on the image. Sections of the vessel wall that are nearly horizontal yield the highest amplitude echoes and hence appear brightest because they were closest to being perpendicular to the ultrasound beams during imaging. Sections where the vessel is slightly inclined are seen as less bright.


Some soft tissue interfaces are better classified as diffuse reflectors. The reflected waves from a diffuse reflector propagate in various directions with respect to the incident beam. Therefore, the amplitude of an echo from a diffuse interface is less dependent on the orientation of the interface with respect to the sound beam than the amplitude detected from a specular reflector.



Scattering


For interfaces whose dimensions are small, reflections are classified as “scattering.” Much of the background information viewed in Figure 2-4 results from scattered echoes, where no one interface can be identified but usually echoes from many small interfaces are picked up simultaneously. The scattered waves spread in all directions, as suggested in Figure 2-6. Consequently, there is little angular dependence on the strength of echoes detected from scatterers. Unlike the vessel wall, which is best visualized when the ultrasound beam is perpendicular to it, the scatterers are detected with relatively uniform average amplitude from all directions. Echoes resulting from scattering within organ parenchyma are clinically important because they provide much of the diagnostic detail seen on ultrasound scans.



In Doppler ultrasound, blood flow is detected by processing signals resulting from scattering by red blood cells. At diagnostic ultrasound frequencies, the size of a red blood cell is very small compared with the ultrasonic wavelength. Scatterers of this size range are called Rayleigh scatterers. The scattered intensity from a distribution of Rayleigh scatterers depends on several factors: (1) the dimensions of the scatterer, with a sharply increasing scattered intensity as the size increases; (2) the number of scatterers present in the beam (e.g., Shung has demonstrated that when the hematocrit is low, scattering from blood is proportional to the hematocrit6); (3) the extent to which the density or elastic properties of the scatterer differ from those of the surrounding material; and (4) the ultrasonic frequency. (For Rayleigh scatterers, the scattered intensity is proportional to the frequency to the fourth power.)



Nonlinear Propagation


A sound wave traveling through tissue will also undergo gradual distortion with distance if the amplitude is high enough. This is a manifestation of nonlinear sound propagation, and it leads to creation of harmonic waves, or waves that have frequencies that are multiples of that of the original transmitted wave. When partial reflection of the distorted beam occurs at an interface, the reflected echo consists of both the original, “fundamental frequency signals” and harmonic components. A 3-MHz fundamental echo is accompanied by a 6-MHz second harmonic echo and so on. Higher-order harmonics are possible, but attenuation in tissue usually limits the ability to detect them. Although the second harmonic echoes themselves are of lower amplitude than the fundamental, it is possible to distinguish them from the fundamental in the processor of an ultrasound machine and to use them to construct an image, called a tissue harmonic image.7


A noteworthy character of tissue harmonic images is that they appear less noisy and have fewer reverberation artifacts than images made with the fundamental. This is believed to be related to the way the harmonic component of the beam forms (i.e., the harmonics gradually grow in amplitude with increasing depth). The harmonic is not present at the skin surface but gradually develops as the beam propagates deeper and deeper into tissue. The second harmonic reaches a peak at some intermediate depth in the patient, then reduces with further increases in depth. Any reverberations or other sources of acoustic noise generated when the transmitted pulse is near the skin surface preferentially contain fundamental frequencies because the harmonics have not built up to any appreciable level at that point. Examples of harmonic images are presented later in this chapter.



B-Mode Imaging



Range Equation


Ultrasound imaging is done using pulse-echo techniques. An ultrasonic transducer is placed in contact with the skin (Figure 2-7). The transducer repeatedly emits brief pulses of sound at a fixed rate, called the pulse repetition frequency, or PRF. After transmitting each pulse, the transducer waits for echoes from interfaces along the sound beam path. Echo signals picked up by the transducer are amplified and processed into a format suitable for display.



The distance to a reflector is determined from the arrival time of its echo. Thus,



image    (2-5)



where d is the depth of the interface, T is the echo arrival time, and c is the speed of sound in the tissue. The factor 2 accounts for the round-trip journey of the sound pulse and echo. Equation 2-5 is called the range equation in ultrasound imaging.8 A speed of sound of 1540 m/sec is assumed in most scanners when calculating and displaying reflector depths from echo arrival times. The corresponding echo arrival time is 13 µs/cm of the distance from the transducer to the reflector.




Forming the Image


Two well-known echo display techniques are also illustrated in the lower two panels of Figure 2-8. The amplitude-mode (A-mode) display is a presentation of the echo signal amplitude versus the echo return time, or the reflector depth. This is a one-dimensional display portraying echo signals and their amplitudes along a single beam line (i.e., along one direction). In contrast, the more versatile B-mode display is used for gray-scale imaging. The display is formed by converting echo signals to dots on a monitor, with the brightness indicating echo amplitudes.


In B-mode scanning, sound beams are swept over a region (Figure 2-9), and echo signals are registered on a two-dimensional (2D) matrix in a position that corresponds to their anatomic origin. Registration is done by placing the B-mode dots along a line that corresponds to the axis of the ultrasound beam as it sweeps across the scanned field; the proper depth of each echo is determined from the arrival time. In Figure 2-9, the sound beam is swept by electronic switching between groups of elements in a linear array transducer. The B-mode display on the monitor follows the axis of the ultrasound beam as it is swept across the imaged region. Usually, 100 to 200 or more separate ultrasound beam lines are used to construct each image. Most ultrasound systems have controls that allow the operator to vary the beam line density, either directly or indirectly when some other image-processing control is manipulated.




Image Memory


An image memory, or scan converter, temporarily retains images for review and photography and converts the image format into one that can be viewed on a video monitor or that can be recorded on videotape. The scan converter is a digital device and may be thought of as a matrix of pixels (image elements); typically, 500 or more pixels are arranged vertically, and about 500 horizontally. The more pixels horizontally and vertically, the better the detail that is represented in the memory, which is particularly important if a postprocessing digital zoom is applied.


Image attributes such as the echo amplitude at each pixel location are represented using a sequence of 1s and 0s, as is the practice for digital devices. The fundamental unit of storage in a digital device is a singular entity called a bit. A single bit can take on a value of either 1 or 0, but by grouping bits into multibit storage cells, each multibit word can represent a large range of values because of the different combinations of 1 and 0 that can be accommodated. For example, “8-bit” memories divide the echo signal into 255 (28) different amplitude levels and store an appropriate level at each pixel location. Twelve-bit memories represent the echo amplitudes using 4096 (212) levels, and so forth. The more bits (amplitude levels), the more different shades of gray are possible from the stored image, especially during postprocessing (see later). Modern scanners also allow storage of cine loops, using a memory that can retain many separate images.


A variety of types of storage media are used in ultrasound. Some laboratories continue their use of hard copy, such as film or other print media. For studies where flow or other dynamic information must be viewed, video tape recorders can store significant quantities of information and facilitate archiving.


Today’s ultrasound machines are equipped with digital storage devices, including fixed computer disks, removable magnetic media such as ZIP disks, and CD-ROMs, and these devices are used to archive study results. Software on the machine can be invoked to recall specific studies and display the image or cine loop sequence. In addition, the majority of installations now utilize computer networks for transferring images, making it possible to view study results on workstations and archive information in centrally organized digital collections. Picture archiving and communication system (PACS) software is available to do these tasks, either on the ultrasound machine itself or off-line. A standard file organization system, the Digital Imaging and Communications in Medicine (DICOM) standard, was created by the National Electrical Manufacturers Association and other standards bodies to aid in the distribution and viewing of ultrasound and other medical images created by equipment from different manufacturers. Each DICOM file contains a “header section” that has information including the patient’s name, the type of scan, image dimensions, and more, as well as the image data itself. Some scanners require a converter box to accept the image data from the scanner, convert it to a DICOM file, and then transfer the file to the PACS network. More commonly, scanning machines themselves have software to convert files to DICOM format and communicate with the external PACS network. When files are in DICOM format, users with access either to the archived data on the scanning machine or from the network itself can employ DICOM readers available for workstations and personal computers to view, archive externally, print, and manipulate the image data.



Frame Rate


In most applications, B-mode imaging is performed with “real-time” scanning machines. These machines automatically sweep ultrasound beams over the imaged region at a rapid rate, say 30 sweeps per second or higher. The image frame rate is the number of complete scans per second carried out by the system. Fundamentally, image frame rates are limited by the sound propagation speed in tissue. An image is produced in the machine by sending ultrasound pulses along 100 to 200 different beam directions (beam lines) into the body. For each beam line, the scanner transmits a pulse and waits for echoes along that beam line, all the way down to the maximum depth setting. Then it transmits a pulse along a new beam direction and repeats the process. Beam lines are addressed serially, meaning the scanner does not transmit a pulse along a new beam line until echoes have been picked up from the maximum depth in the previous line. The speed with which the pulse propagates through tissue, the depth setting of the scanner, the number of transmit focal zones, and the number of beam lines used to form a single image frame all intermix to establish the maximal possible image frame rate.


Using the range equation, if the maximum depth setting is D, it takes a time (T = 2D/c) to receive echoes from the entire beam line. The amount of time for a complete image frame constructed with data from N beam lines is simply N × T, or 2 ND/c. If the maximum frame rate is FRmax, FRmax will be equal to the inverse of the time needed for a complete image. This may be written as



image    (2-6)



For soft tissue in which the speed of sound is about 1540 m/sec, or 154,000 cm/sec, if the depth setting (D) is expressed in centimeters, Equation 2-6 also works out to



image    (2-7)



For example, with N = 200 beam lines and an image depth of 15.4 cm, FRmax is 25 Hz.


Operators can easily verify that reducing the depth setting on the machine will increase the frame rate, and vice versa. Often, the machine is programmed to provide as high a frame rate as is practical for the operator settings. Some machines allow the operator to change N, the number of beam lines used to form the image, for example, by increasing the angular separation between beam lines. This, in turn, also affects the frame rate, as does changing the horizontal size of the image and changing the number of transmit focal zones.



Transducer Properties


An ultrasound transducer provides the communicating link between the imaging system and the patient. Medical ultrasound transducers use piezoelectric ceramic elements to generate and detect sound waves. Piezoelectric materials convert electric signals into mechanical vibrations and pressure waves into electric signals. The elements, therefore, serve a dual role of pulse transmission and echo detection.


Internal components of an array transducer are shown in Figure 2-10. In the figure, the elements are seen from the side, and the ultrasound waves would be projected upward. The thickness of the piezoelectric element governs the resonance frequency of the transducer. Quarter-wave matching layers between the piezoelectric elements and a protective outer faceplate are used on most transducers. Analogous to special optical coatings on lenses and on picture frame glass, the matching layers improve sound transmission between the transducer and the patient. This improves the transducer’s sensitivity to weak echoes. Backing material is often used in pulse-echo applications to dampen the element vibrations after the transducer is excited with an electric impulse. Dampening shortens the duration of the transmitted pulse, improving the axial (or range) resolution. With optimized designs of the matching and backing layers, transducers can be made to operate over a range of frequencies. Hence, ultrasound machines provide a frequency control switch that the operator manipulates to select the frequency from a menu of choices available for each probe. Some transducers have sufficient frequency range that harmonic imaging can be done, where a low-frequency transmit pulse is sent out, and echoes whose frequency is twice that transmitted are detected and used in imaging.



Mar 5, 2016 | Posted by in ULTRASONOGRAPHY | Comments Off on Physics and Instrumentation in Doppler and B-mode Ultrasonography

Full access? Get Clinical Tree

Get Clinical Tree app for offline access