Optimizing the scan

7 Optimizing the scan




INTRODUCTION


The preceding chapters have covered some of the basic scientific principles behind ultrasound, the Doppler effect, and hemodynamics. The sonographer should now have a clearer understanding of how a B-mode image is created and how color flow imaging can be used to interrogate blood flow in vessels rapidly, allowing the blood flow in selected areas to be assessed with spectral Doppler. This has made duplex scanning a powerful technique for the investigation of patients with vascular diseases, and many vascular surgeons are making clinical decisions on the basis of duplex scanning alone. It is therefore vitally important that the operator understands the use of the scanner controls and the limitations of the technique. Some manufacturers have introduced auto-optimization controls, but there are still many situations in which the controls will need to be adjusted manually. Manufacturers often use different names or terms for the same scanner control or function, such as power imaging and color angiography, both of which relate to power Doppler imaging. Another interchangeable control used by manufacturers is pulse repetition frequency (PRF) and scale. It is important to consult the operator’s manual or ask the manufacturer if the function of any control is not clear.


Ultrasound scanners have a range of examination-specific presets that optimize the system for a particular examination and it is important to start the scan with the appropriate preset selected. However, in many instances the scanner controls need adjusting, or optimizing, to demonstrate pathology. In addition, a number of imaging and Doppler artifacts may be confused or misinterpreted as significant disease, leading to serious diagnostic errors. The aim of this chapter is to introduce the sonographer to the practical aspects of scanning, covering the basic use of scanner controls and reiterating some of the principles discussed in the preceding chapters. Imaging artifacts will also be discussed to assist the sonographer in the interpretation of images. It is likewise essential that the sonographer has a good understanding of the principles relating to ultrasound safety in order to minimize any exposure risks to the patient.




STARTING THE SCAN






The scan should be carried out in a dimly lit room to optimize visualization of the black and white image. The transducer selected should be of the highest frequency that allows adequate penetration to the area to be examined. When scanning, it is important to adopt a logical approach. Using a systematic technique cuts down on examination time and ensures that pathology is less likely to be missed. The scan is best started by examining the region of interest with B-mode imaging alone, to identify relevant structures. Avoid switching on the color flow or spectral Doppler straight away, unless they are essential for identifying vessels, as the imaging frame rate will be reduced and the display may be confusing if anatomy has not been clearly identified.




IMAGING ARTIFACTS


An imaging artifact is a feature on the image that does not relate exactly to a structure within the tissue being investigated. This can be due to a feature being misplaced on the image, a feature appearing that is not present within the tissue, or an existing structure that is absent from the image. The creation of an image relies on the assumption that the ultrasound beam travels in a straight path between the transducer and the structures within the tissue and returns along the same path once reflected. It is also assumed that the attenuation of tissue is constant. Any process that alters this situation can lead to the misplacement or absence of information. This can be caused by the following:



Multiple reflections can lead to reverberation artifacts, seen as several equidistant echoes that reduce in brightness with depth. This is due to multiple reflections, along the same path, between the transducer and a strongly reflecting boundary (Fig. 7.1A) or between two parallel, strongly reflecting surfaces (Fig. 7.1B). If the multiple reflections do not return along the same path, the structure may be misplaced on the image (Fig. 7.1C).


A mirror image of a structure can be produced in the presence of a strongly reflecting surface. Figure 7.1D shows how the true position of a structure is displayed, with a second ghost image also displayed. The ghost image has been created by an ultrasound beam that has undergone multiple reflections from the strongly reflecting surface.


Refraction can lead to bending of the path of the ultrasound when the beam passes through an interface between two media in which the speed of sound is significantly different (see Fig. 2.7).


Range ambiguity can occur if an echo from the previously transmitted pulse is received back from a distant boundary after the current ultrasound pulse has been transmitted. The scanner will assume that the echo is from the current pulse and place it nearer to the top of the image rather than at its true depth.


Grating lobes are areas of lower-intensity ultrasound outside the main beam and are produced as a function of the multi-element structure of array transducers. These grating lobes can lead to strongly reflecting surfaces outside the main beam being displayed in the image.



The image in Figure 7.2

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Dec 26, 2015 | Posted by in CARDIOVASCULAR IMAGING | Comments Off on Optimizing the scan

Full access? Get Clinical Tree

Get Clinical Tree app for offline access