Imaging Principles in Magnetic Resonance Imaging




Magnetic resonance imaging (MRI) is based on spatially encoding a nuclear magnetic resonance (NMR) signal. Therefore an understanding of the basics of NMR is critical for understanding MRI, and this information is given in the first section of this chapter. Magnetic field gradients are used in MRI to spatially encode the NMR signal to obtain the raw MRI data, or k-space. Images are reconstructed from k-space using mathematical algorithms, most commonly the Fourier transform. Image contrast is controlled by changing the timings of the encoding using specific pulse sequences. Thus in addition to a basic explanation of NMR signal generation, the process of spatial encoding and contrast generation are described. Additionally, commonly used rapid imaging techniques and frequently encountered artifacts are introduced.


Physics Principles


Nuclear Spin


All atoms are made up of electrons surrounding an atomic nucleus, which can contain protons and neutrons. The nucleus of an atom can have a quantum property known as nuclear spin. For instance, a hydrogen atom has a proton as its nucleus, and the magnetization from a macroscopic sample of proton spins is what is commonly measured in NMR and MRI. Many other atoms have nuclei with this property, but the discussion here will focus on hydrogen nuclei (i.e., protons) because hydrogen is a key component of biological tissue (e.g., water, fat) and most commonly visualized in clinical MRI.


The behavior of each individual nucleus is described by quantum mechanics. However, when working with many nuclei in bulk, the picture can be simplified to make it more easily understandable. In this treatment the magnetization from a single proton can be thought of as an arrow pointing in a specific direction in three-dimensional (3D) x-y-z space. In the absence of an externally applied magnetic field, the nuclear magnetization is randomly oriented; in other words the magnetization from each proton in a group of protons (i.e., in a cup of water or the human body) is not aligned. If one were to try to measure the net magnetization or aggregate magnetization coming from the group of protons, it would be zero because each arrow would be pointing in a different direction, resulting in total cancellation.


However, in the presence of an external magnetic field such as that used in NMR or MRI, the nuclei act like tiny bar magnets and align themselves with the field in one of two states—parallel or antiparallel. If an equal number of protons were in each state, the net magnetization would again be zero (the magnetic field from the nuclei pointing up would cancel the field from the nuclei pointing down). However, the state parallel to the external magnetic field has a slightly lower energy level than the antiparallel state, so slightly more nuclei end up in the parallel state than in the antiparallel state. This small difference in parallel and antiparallel protons leads to the signal used for imaging in MRI. For clinical MRI field strengths, only one or two protons in a million can be detected; this number increases as the MRI field strength increases (which leads to an increase in the signal-to-noise ratio [SNR] at higher field strengths). The net magnetization, or magnetization from these excess protons, can be visualized as a larger arrow pointing in the direction of the external magnetic field, where the size of the arrow indicates the magnitude of the magnetization.


Nuclear Magnetic Resonance


The net magnetization of protons in an external magnetic field aligns with the external magnetic field. This is called the equilibrium state of the magnetization and is usually defined as the z-direction ( Fig. 3-1A [red arrow] ). However, by applying energy in the form of electromagnetic waves at precisely the right frequency, the net magnetization can be made to tip away from the equilibrium position into the x-y plane. This frequency is known as the Larmor frequency:


<SPAN role=presentation tabIndex=0 id=MathJax-Element-1-Frame class=MathJax style="POSITION: relative" data-mathml='ω=γB0′>?=??0ω=γB0
ω = γ B 0
where ω is the precession frequency, B 0 is the strength of the external magnetic field, and γ is a constant that differs for each atomic nucleus, known as the gyromagnetic ratio ). For instance, protons (with a gyromagnetic ratio of 42.577 MHz/T) in a magnetic field of 3.0 tesla (T) will be tipped when energy with a frequency of 127.731 MHz is applied. This energy is transmitted by so-called transmit coils and is usually referred to as a radiofrequency (RF) pulse. The angle by which the magnetization is tipped is called the tip angle. Once displaced from equilibrium, the net magnetization will precess around the external magnetic field. In other words, if the magnetization is tipped away from the z-axis to the x-axis using a 90-degree RF pulse (see Fig. 3-1A [orange arrow] ), the end of the arrow describing the magnetization will move around the x-y plane, tracing a circle (yellow, green, and blue arrows). The frequency of this precession also depends directly on the external magnetic field according to the Larmor relationship shown above: the stronger the field, the faster the magnetization will precess.


FIG 3-1


A, Magnetization depicted on a 3D axis. Before excitation the magnetization is aligned along the z-axis (red arrow). After a 90-degree RF pulse the magnetization is tipped into the x-y plane. Here it is shown as being aligned with the x-axis (orange arrow). Once in the x-y plane the magnetization begins to precess in the x-y plane, first moving between the x and y axes (yellow arrow) and then aligning with the y-axis (green arrow) and after a longer time pointing in the negative direction along the x-axis (blue arrow). The magnetization will eventually precess around to the x-axis again and continue to move around in the x-y plane. Note that the effect of a 180-degree pulse would be to completely invert the magnetization (purple arrow). B, The signal collected from a receiver coil along the x-direction. Immediately after the RF pulse, the signal along the x-direction is maximal, leading to a large signal (orange dot). As the magnetization precesses, the signal decreases (yellow dot) until it reaches a minimum when the magnetization points along the y-direction (green dot). The magnetization continues to precess, leading to a negative signal when it is pointed in the negative x-direction (blue dot). Continued precession leads to the sinusoidal signal shape over time.


Signal can only be picked up from magnetization that is precessing in the x-y plane. An RF receiver coil is used to collect the signal after it is tipped into this plane using an RF pulse. In simple terms the receiver coil “sees” magnetization along one direction; in other words, voltage is induced in the receiver coil because of the change in direction of the magnetization vector that results from the precession (similar to an electrical generator). To demonstrate the relationship between the precession of the magnetization and the signal, the x-direction will be used as an example here. When the magnetization is pointed toward the x-axis, the signal will be at a maximum (see Fig. 3-1B [orange dot] ). When the magnetization precesses to be completely on the y-axis, there is no magnetization in the x-direction, so the signal picked up by the receiver coil is zero (see Fig. 3-1B [green dot] ). When the magnetization precesses further into the −x direction, the signal will have a minimum (see Fig. 3-1B [blue dot] ). The signal over time seen by precessing magnetization traces out a sinusoid, which can be appreciated in Figure 3-1 . This treatment ignores relaxation, which is discussed next.


Relaxation


The net magnetization in an external static magnetic field aligns with the magnetic field, but an RF pulse can be used to tip this magnetization into the x-y plane as explained earlier. Once tipped and with no further perturbation, the magnetization returns to the equilibrium state along the z-axis. This return to the equilibrium state is known as T1 relaxation, otherwise known as spin-lattice relaxation or longitudinal relaxation. T1 relaxation follows an exponential recovery curve, as shown in Figure 3-2A , which is described by the following equation (assuming the signal started in the x-y plane after a 90-degree RF pulse):


<SPAN role=presentation tabIndex=0 id=MathJax-Element-2-Frame class=MathJax style="POSITION: relative" data-mathml='s(t)=1−e−t/T1′>?(?)=1??/?1s(t)=1−e−t/T1
s ( t ) = 1 − e − t / T 1
where the signal at time t is written as s(t). Different tissues have different T1 times based on properties including the physical state of the components of the tissue (liquid, fibrous, solid) and the mobility of the protons in the tissue. Two example T1 relaxation curves for tissues with different T1 values are shown in Figure 3-2A . The tissue with the shorter T1 requires less time for the protons to return to the equilibrium state (i.e., aligned with the z-axis), whereas the tissue with a longer T1 time relaxes more slowly (see Fig. 3-2A ). These differences in T1 times can be used to generate contrast in MRI images. By collecting data when the differences between the relaxation curves is largest, a maximum T1-weighting between two tissues can be achieved. To increase these differences in T1 and thus increase the contrast between tissues, an external contrast agent can be injected to shorten the T1 of one tissue. Gadolinium-based contrast agents are often injected into the blood pool to shorten the T1 time of blood and/or perfused tissues, which (given appropriate pulse sequence settings) can cause a large difference between signals from tissues in the proximity of the contrast agent and other tissues unaffected by the contrast agent.


FIG 3-2


A, A plot showing the T1 relaxation of two tissues with different T1 values after a 90-degree RF pulse. Immediately after the pulse the longitudinal magnetization (Mz) is zero because all of the magnetization points along the x-y plane. Over time the magnetization slowly returns to equilibrium. The curve shown in black describes the relaxation of a tissue with a long T1 value (1000 ms); note that it takes a long time for the magnetization to return back to the equilibrium value. The red curve describes the relaxation of a tissue with a short T1 value (200 ms); this curve shows that the longitudinal magnetization is restored more quickly. B, A plot showing the T2 decay for two tissues with different T2 values. Immediately after a 90-degree RF pulse, all of the magnetization is found in the x-y plane. As the spins precess, they start to dephase, leading to a decrease in the net magnetization, or T2 decay. The red curve shows the signal time course for a tissue with a long T2 value (200 ms), and the black curve shows a tissue with a short T2 value. The signal coming from the tissue with the shorter T2 decays faster, and this effect is used to generate T2 contrast in MRIs.


In addition to T1 relaxation, a second relaxation mechanism is often used to impart contrast to MRI. This second relaxation parameter is T2 relaxation, otherwise known as spin-spin or transverse relaxation. The T2 parameter describes the behavior of the magnetization as it precesses in the x-y plane. Recall that the net magnetization is actually a sum of the magnetization of many protons. If all the protons precess at exactly the same frequency, the net magnetization would retain the same magnitude as it precesses. However, if some of these spins precess at slightly higher frequencies (i.e., a bit faster) and some at slightly lower frequencies (i.e., a bit slower), the magnetization will begin to “fan out” in the x-y plane. The net sum of this fanned out magnetization will have a smaller magnitude than when all the spins precess at exactly the same frequency. Over time the slight differences in precession frequency will lead to magnetization that is spread in all directions in the x-y plane, leading to a net magnetization of 0. This T2 relaxation process is described using the following equation:


<SPAN role=presentation tabIndex=0 id=MathJax-Element-3-Frame class=MathJax style="POSITION: relative" data-mathml='s(t)=1−e−t/T2′>?(?)=1??/T2s(t)=1−e−t/T2
s ( t ) = 1 − e − t / T2


The smaller the T2 value, the faster the magnetization fans out around the x-y plane, leading to a rapid decrease in the signal (see Fig. 3-2B ). Like T1, different tissues have different T2 values depending on the makeup of the tissue. Small disturbances in the local magnetic field of the tissue can cause rapid T2 decay as well as differences in the chemical environments. In a broad sense, T2 provides information about how fluidlike a particular tissue is; solid tissues have shorter T2 values, whereas pure liquid water has a very long T2 value. It is important to note that the T1 and T2 times of tissues are not mathematically related, and MRIs are often weighted by one (or both) of these parameters to highlight desired structures. However, T2 is always shorter than T1.


Although the T2 value is an intrinsic tissue property, the transverse magnetization can also be affected by small differences in the magnetic field. Any inhomogeneities in the field will cause individual nuclei to have slightly different resonance frequencies and lead to a more rapid fanning out of the magnetization in the x-y plane. The parameter that describes the relaxation of magnetization in the transverse plane secondary to field inhomogeneities is T2*. The actual T2* depends on the local magnetic field, which can be affected by differences in magnetic susceptibility of adjacent materials (i.e., tissue-air interfaces) or the presence of iron. T2* is always shorter than T2, and although T2 relaxation cannot be avoided, the effects of T2* relaxation can be eliminated by using some specialized imaging techniques (e.g., spin echo sequences).


Chemical Shift


So far it has been assumed that all protons have the same precession frequency, also called the resonance frequency. However, the Larmor equation actually only describes the resonance frequency of an isolated hydrogen nucleus. The proton is almost never found completely isolated; hydrogen is normally bound to another nucleus (in water, this is oxygen) through chemical bonds. These chemical bonds are made up of electrons moving around the nuclei of the bound atoms, and this cloud of electrons serves to “shield” the protons of the hydrogen atom from the externally applied main magnetic field. Thus the actual magnetic field seen by the protons is usually slightly less than the actual applied field, as described in the following equation:


<SPAN role=presentation tabIndex=0 id=MathJax-Element-4-Frame class=MathJax style="POSITION: relative" data-mathml='ωc=γ⋅(1−σ)B’>??=?(1?)?ωc=γ⋅(1−σ)B
ω c = γ ⋅ ( 1 − σ ) B
where ω c describes the true resonance frequency of the proton and σ describes the degree of shielding from the electrons. The value of σ is determined by the types of chemical bonds; for instance, the degree of shielding, and thus the resonance frequency, is different for water and fat. This effect, called chemical shift, is exploited for NMR spectroscopy, a technique used to characterize different chemical compounds. Chemical shift is also observed in MRI, and this effect will be described later in the chapter.


The Basic NMR Experiment


The basic concepts described can be put together to understand how a simple hydrogen NMR experiment works. The timing of the events in the experiment is described using a pulse sequence diagram, where each line represents a different type of event. The basic NMR pulse sequence diagram in Figure 3-3 shows when RF pulses are applied (top) and when data are collected (bottom).




FIG 3-3


A basic NMR pulse sequence. Both lines describe the timing of events in the pulse sequence. The top line shows the RF pulse. The bottom line shows when signal is collected (ACQ). Note that the signal shape is an exponentially decaying sinusoid owing to relaxation.


A sample made up of substances containing protons is placed in an NMR scanner with a strong external magnetic field, B 0 . This field causes the protons to align, leading to a net magnetization. An RF pulse is applied to tip the magnetization from the equilibrium z-axis into the x-y plane. Once in the x-y plane the magnetization begins to precess at a frequency determined by the chemical environment of the hydrogen nuclei. Signal from this precessing magnetization is collected using a receiver coil. During signal collection the magnetization experiences both T1 and T2 relaxation simultaneously. The precession causes the signal to oscillate, and the relaxation causes the signal to decay to zero, leading to an overall signal form that looks like an exponentially decaying sinusoid (this shape is shown on the data acquisition line labeled “ACQ” in Fig. 3-3 ). This process can be repeated several times and the signal averaged to increase the SNR (see “ Imaging Time and Performance Tradeoffs ”).


Such an experiment can be used to determine how many and what type of chemical compounds are present in a sample by examining the resonance frequencies in the collected signal. However, because the receiver coil picks up signal from all parts of the sample, there is no information about the spatial distribution of the chemical compounds in the sample.




Imaging


Introduction


NMR is traditionally used to determine the chemical structure of a sample and does not require spatial localization. However, MRI depends on being able to distinguish between spins at different physical locations, so spatial localization is performed by applying magnetic field gradients in each of the three spatial axes in addition to the main static magnetic field. A schematic of this idea is shown for one-dimensional (1D) encoding in Figure 3-4 . Although this picture does not fully explain how two-dimensional (2D) or 3D spatial localization is performed, this idea forms the basis for spatial encoding in MRI, which will be explained in detail in the following sections.




FIG 3-4


A, Three vials are filled with a liquid (red) and placed in an MRI scanner with a main magnetic field strength of B 0 . A magnetic field gradient, G x , which varies along the x-direction, is also applied; the net magnetic field at any position in space depends on the combined strength of both of these magnetic fields. After an RF pulse the magnetization in the left vial precesses slowly, the center vial moderately, and the right vial rapidly because of the different magnetic fields they experience. If signals in time were collected from each vial independently, these differences in precession rate would be seen, and differences in the amplitude of the signals would correspond to the amount of liquid at each of the spatial positions (middle). The summed signal (bottom) is what is actually collected by the MRI scanner. This type of encoding, where a gradient is played out during signal acquisition, is called frequency encoding. B, The summed MRI signal can be decomposed into separate frequencies using the Fourier transform. The time signal is converted into frequency units, which reflects the position of the vials. This can be represented as a 1D “image” where frequencies with high signal amplitude are bright and frequencies with low signal amplitude are dark.


In Figure 3-4A , three signal sources (liquid in tubes) are shown in red. Here a linearly varying magnetic field is applied in the x-direction; this field gradient is called G x . At the center the amplitude of this additional magnetic field is 0, and the spins at that location precess at the Larmor frequency determined by the main magnetic field. However, the spins on the left side experience a total magnetic field that is slightly less than the main magnetic field owing to the gradient. These spins will precess with a slightly lower frequency. Similarly the spins on the right side experience a slightly higher magnetic field and precess at a slightly higher frequency. Thus the result of applying this magnetic field gradient is to differentiate spins at different locations by causing them to precess at different frequencies. In this way a one-to-one mapping can be made that links the location of the spins to their frequency. Additionally, note that the left-most vial contains less fluid and the signal coming from this vial has a lower amplitude than the center vial, which also has a lower amplitude than the right-most vial. In fact the signal amplitude at each frequency is directly proportional to the number of spins precessing at that frequency. The signals coming from the different locations, however, cannot be separated by the receiver coil. Only the sum of the three signals is actually collected. This signal is shown at the bottom of Figure 3-4A , and the next section describes how image information is derived from this composite signal.


Data to Image: Fourier Transform


The schematic shown in Figure 3-4A and the legend explain how spatial encoding is performed but do not explain how spatial information is extracted from the signal to form an image in MRI. As described earlier the MRI signal is collected from the entire volume using a receiver coil. Thus the challenge is to separate individual single frequency signals from the complex multifrequency data. Mathematics has a simple way to perform this separation: the Fourier transform, which decomposes a signal in time into its component frequencies—exactly the problem that must be solved in MRI. By finding the Fourier transform of the MRI signal, a “map” of the frequencies contained by the signal can be derived. Because the frequencies correspond directly to the spatial locations of the spins, this frequency map is also equivalent to the 1D “image” of the spins. Note that in the 1D image shown in Figure 3-4B , the vial on the left with little liquid has a low signal intensity, the center vial a medium intensity, and the vial on the right a high intensity. All frequencies that were not present in the signal correspond to zeros in the frequency map and black in the 1D image.


Frequency Encoding


Nearly all MRI scans perform spatial localization in one dimension using frequency encoding, which is the closest form of encoding to the basic schematic shown in Figure 3-4 . In frequency (or read) encoding, a magnetic field gradient is applied to the object in one direction while data are collected (and hence the alternative name, read encoding ). The spins precess at a frequency dictated by their local magnetic field, and the resulting data are a complex combination of all of the different frequencies, as described earlier. If a 1D Fourier transform is applied to these data, a 1D projection through the object would result, as shown at the bottom of Figure 3-4B . Only one dimension can be resolved using frequency encoding, because the second dimension is not encoded and the spins in that direction are not distinguishable from one another. For this reason, another encoding mechanism must be employed in addition to frequency encoding, namely phase encoding.


Phase Encoding


Phase encoding works through the same basic mechanism as frequency encoding. The only difference is that the magnetic field gradient used in phase encoding is not applied while the data are being collected. Instead a phase-encoding gradient is turned on in a direction orthogonal to the readout direction before data collection. While the phase-encoding gradient is turned on, the spins precess with a frequency that depends on their spatial location ( Fig. 3-5A ). Once the gradient is turned off the spins will stop precessing, but they will retain the phase they had due to the phase-encoding gradient (hence the name). A readout gradient is then applied in the second encoding direction, as described earlier, and data are collected.




FIG 3-5


A, A phase-encoding gradient is applied along the y-direction, which causes spins in different locations in the y-direction to precess with different frequencies. When the gradient is turned off, the spins have a phase that is dependent on their position. This phase is retained when frequency encoding is applied in a direction orthogonal to the phase encoding. B, To collect different phase-encoding lines, phase-encoding gradients with different strengths are used. For k y = 0, no phase-encoding gradient is applied, and no y-position–dependent phase is imparted. For k y = 1, a gradient is applied with a strength and duration that leads to one period of a sinusoid across the field of view (FOV) in the y-direction. For k y = 2, the gradient area is twice that for k y = 1, and the phase variation of the magnetization over the FOV covers two periods.


The application of one phase-encoding gradient would not help to make a 2D image, because just imparting a phase is not enough to spatially encode information. Thus many phase-encoding gradients with different strengths are applied, followed by a readout after each. The effect of applying phase-encoding gradients with different strengths is shown in Figure 3-5B . Here, schematic gradients are shown for k y = 0, k y = 1, and k y = 2. Note that the area of the gradient (time multiplied by amplitude) determines how much phase is accumulated across the field of view (FOV). If each readout line for different phase-encoding gradient magnitudes are collected and put into a 2D grid, the resulting data matrix is known as k-space.


k-Space


k-Space is the commonly used name for the raw data collected in MRI. These data are a set of sampled points stored in a matrix of numbers, where each line of numbers corresponds to one value of the phase-encoding gradient. The units of k-space are time (s) or inverse meters (m −1 ); both are different ways of looking at the same data. The k-space data are encoded as previously described, with phase and frequency encoding using magnetic field gradients. The gradients control the path that is traversed through k-space. An example of how gradients are used to “drive” through k-space is shown in Figure 3-6 . The phase-encoding gradient is used to move up and down in k-space, and the readout gradient is used to move to the left and right.




FIG 3-6


A, Schematic of an MRI pulse sequence, where the RF pulse and slice select gradients are shown in blue, the phase-encoding gradients in yellow and orange, the read dephasing gradient in green, the frequency encoding gradient in pink, and the acquisition window in gray. B, Each gradient is used to “move” around in k-space. After slice selection, no spatial encoding has yet been performed, and the location in k-space is denoted with the blue dot (center). The phase-encoding gradient enables motion up and down in k-space (orange and yellow arrows), the read dephasing gradient causes motion to the left (green arrow) , and the frequency encoding gradient to the right (red arrow). If data are collected during the frequency encoding gradient, the result is a line of k-space data.


Once the k-space data have been collected, a discrete Fourier transform is applied to mathematically transform the data to the image domain. The Fourier transform takes the complex multifrequency data and reports back which frequencies are present in the data. The result is a matrix of numbers that correspond to the image. The units of the image are the inverse of the k-space data, namely inverse seconds (s −1 , or Hz) or meters (m), which matches with the idea that the image is a map of the frequencies generated using phase and frequency encoding as previously described.


Resolution and Field of View


The FOV is defined as the total size of the image (in units of distance). The matrix size of the image is often a power of 2 (128, 256, 512) owing to historical restrictions on performing a fast Fourier transform. Given the FOV and matrix size for an image, resolution is defined as the FOV divided by the matrix size, resulting in the nominal pixel or voxel size. For instance, if a 2D image has an FOV of 200 mm × 200 mm with a matrix size of 128 × 96 in the read and phase-encoding directions, the spatial resolution (pixel size) would be 1.56 mm × 2.08 mm in these directions.


The FOV and resolution of the image are directly related to the spacing and coverage of the data collected in k-space. The spacing between adjacent points in k-space determines the FOV in the image. The closer the phase-encoding lines or readout points are in k-space, the larger the FOV will be. In general the k-space spacing is chosen to be just large enough to cover the entire object. If the spacing between k-space lines becomes too large, the FOV may no longer cover the object, which can lead to aliasing artifacts (described in the “ Parallel Imaging ” and “ Clinical Artifacts ” sections).


The resolution of the image is determined by the extent of the k-space covered. The extent can be described by looking at the product of the spacing between k-space points (which determines the FOV) and the number of points collected. The larger this product is, the higher the image resolution will be. This relationship can be understood by thinking of the image as a frequency map. When more time-domain data (k-space) are collected, more frequencies can be resolved from that data, and the resolution of the image will be higher.


The significance of the different parts of k-space can be examined by removing some portions of data and performing a Fourier transform to see what happens in the resulting image. It is often said that the center of k-space determines the contrast of the image and the outer portions of k-space provide resolution. Figure 3-7 explores this statement by manipulating the k-space. At the top, an example k-space and its corresponding image is shown. If the extent of the k-space is reduced by eliminating some of the outer k-space lines, the resulting image has a lower spatial resolution (it appears blurrier [middle] ). If only the outer portions of k-space are retained and the central portions set to zero, only the edges appear in the image (bottom). The effects of eliminating other areas of k-space are examined later in “Fast Imaging for Clinical Practice.”




FIG 3-7


Top, An example of fully sampled k-space (left) and the corresponding image (right). Middle, If only the center portion of the k-space is collected, the resulting image has a lower resolution but shows the overall image contrast. Bottom, The outer portions of k-space correspond to the high-resolution details of the image.


2D vs. 3D: Slice and Volumetric Encoding


Most clinical MRI images are collected in 2D, where the two spatial dimensions are encoded using read and phase encoding. However, before these dimensions can be encoded the desired 2D slice must be selected from the 3D volume of the body. This process is performed using slice-selection encoding.


Simultaneously with the RF excitation pulse, a slice-selection magnetic field gradient is applied to the subject in the slice direction (perpendicular to read and phase-encoding direction). The RF pulse will only excite magnetization resonating at specific frequencies. Because the resonance frequency can be controlled by application of the slice-select gradient, by setting the pulse sequence parameters correctly, only magnetization inside the desired slice will be excited by the RF pulse, and the magnetization outside the slice will remain mostly untouched. This process is shown schematically in Figure 3-8 . The excited magnetization will then be subjected to read and phase encoding to resolve the spatial information in the 2D slice.




FIG 3-8


A schematic of slice selection using using a magnetic field gradient. A gradient (G ss shown as the orange-to-green wedges) is applied along the direction of the slice to be selected (in this diagram, the x-direction). The total strength of the magnetic field is shown on the vertical axis. If an RF flip pulse is used that excites frequencies between ω 1 and ω 2 but not the frequencies outside of this range, only spins along the x-direction that are precessing at these frequencies will be selected. Thus in this diagram all spins between X 1 and X 2 would then be excited in the ideal case (the selected slice is shown in light gray). Note that if the strength of the slice select gradient is increased (where the new slice select gradient is shown as the light blue wedges), the same RF flip pulse exciting frequencies between ω 1 and ω 2 will result in a slice that is twice as thin as the original slice (shown in dark gray) because only these spins will be precessing at the appropriate frequencies with the increased magnetic field gradient.


As in read and phase encoding, a higher definition in the slice direction (or thinner slice) requires a stronger gradient in the slice direction. If the slice gradient is very weak, the slice will be very thick (light gray slice in Fig. 3-8 ). Such a thick slice can also be selected and then further encoded using a second, independent phase-encoding gradient in the slice direction. Such encoding is termed 3D encoding, and the second phase-encoding operation is often called partition encoding (to distinguish a 2D slice from a single 3D partition); 3D encoding results in a true 3D k-space, requiring a 3D Fourier transform to reconstruct the 3D image. Although 3D encoding is more time consuming, the image SNR is significantly higher because of the larger area covered. However, because the scan time is longer, there is a greater chance for image artifacts due to motion, which is why 3D imaging is used clinically less frequently than 2D imaging.


Pulse Sequences for MRI


Pulse sequences are methods for manipulating the magnetization of tissues and the signals they produce to generate a specific image contrast. An MRI system programmer uses pulse sequences to exploit the NMR principles that produce signal from tissue (and thus contrast among different tissues) and to perform spatial encoding.


The pulse sequence is a series of events triggered over time by the MRI scanner to change the magnetization of tissue in some way. There are two basic ways of manipulating the magnetization of tissue: applying RF energy and applying magnetic field gradients (G x , G y , and G z ). The pulse sequence diagram is a schematic that describes the order and duration of RF and gradient fields being switched on and off over time to produce the type of image contrast weighting and spatial encoding the MRI user wants. Each family of pulse sequences has strengths and weaknesses for different clinical imaging applications.


As described earlier the RF is switched on for a very short time with variable power and then switched off, and this short transmission is called an RF pulse. When it is used to excite the tissue signal for imaging, it is often called an excitation or flip pulse. For a specific MR scan the pulse will be associated with a flip angle that corresponds with the resulting magnetization of the excited tissue (0 degrees is magnetization parallel to the B 0 field, shown by the red arrow in Figure 3-1 , 90 degrees is perpendicular to B 0 , shown by the orange arrow, and 180 degrees is antiparallel to B 0 , shown by the purple arrow). Additional RF pulses may be applied at different times, depending on the type of pulse sequence.


Pulse Sequence Diagrams.


The pulse sequence diagram is an illustration of the time relationship of the different events that make up the pulse sequence. It is not always drawn to scale. There are some common symbols used in pulse sequence diagrams, but there is no standard format. Events in the pulse sequence are represented by symbols located on the respective line at the relative time of the event.


There are five essential system hardware functions whose events are usually shown on a pulse sequence diagram: RF, G slice , G phase , G read , and ACQ ( Fig. 3-9 ). Note that the pulse sequence shown in this figure is a gradient echo sequence, but the same basic elements appear in all pulse sequences. RF pulses are shown on an RF line to indicate any excitation and refocusing pulses. The three orthogonal magnetic field gradients, namely phase, slice, phase, and frequency (or read) encoding, are shown as separate lines. The activation of each of the gradients is indicated by a step change on the appropriate gradient’s line. The activation of the receiver and the data acquisition period (ACQ) is normally shown in context with the other elements of the pulse sequence. Sometimes the acquisition is shown in the same line as the RF pulses because this also represents an RF event.




FIG 3-9


Gradient echo pulse sequence diagram showing symbols used for RF and gradient pulses.

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Aug 20, 2019 | Posted by in MAGNETIC RESONANCE IMAGING | Comments Off on Imaging Principles in Magnetic Resonance Imaging

Full access? Get Clinical Tree

Get Clinical Tree app for offline access