Intraoperative Indocyanine Green Fluorescent Imaging for Prevention of Bile Leakage After Hepatic Resection



Fig. 35.1
Type A pattern of fluorescence (no fluorescence type). Fluorescent imaging showed no fluorescence detected on the cut surface of the liver, suggesting absence of bile ducts at the surgical margin. The cut surface of the liver under conventional white light illumination (a) and fluorescent imaging (b)



A319731_1_En_35_Fig2_HTML.gif


Fig. 35.2
Type B pattern of fluorescence (intact bile duct type). Fluorescent imaging showed one fluorescent duct on the cut surface of the liver (arrow). Common bile duct (arrowhead). There was an intact bile duct corresponding to the fluorescent area on the cut surface (arrow) under conventional white light illumination (a) and fluorescent imaging (b)


Postoperative bile leakage was diagnosed by the following findings: detection of bile leakage from the wound or through the drainage tube (total bilirubin level in the drainage fluid >3 times the serum level), intra-abdominal accumulation of bile confirmed by drainage, or demonstration of bile leakage on postoperative cholangiography.




35.3 Results


Thirty-two of the 132 patients (24 %) had major leakage of ICG, defined as leakage of green dye without fluorescent imaging. Repair was performed by z-suturing with 6–0 nonabsorbable sutures in 28 patients and by ligation with 4–0 absorbable sutures in four patients. The leaks were all successfully repaired during surgery and were further tested with fluorescent imaging.


35.3.1 Patterns of Fluorescence


In 37 of the 132 patients (19 %), no fluorescence was detected on the cut surface of the remnant liver on ICG fluorescent cholangiography, suggesting absence of bile ducts at the surgical margin (type A pattern of fluorescence) (Fig. 35.1). In the remaining 95 patients, the pattern of fluorescence was classified into the following three types: intact bile duct type (type B, fluorescence showed one or more intact bile ducts on the cut surface of the liver; n = 51) (Fig. 35.2), injured bile duct type (type C, leakage of dye from one or more bile duct stumps on the cut surface; n = 31) (Fig. 35.3), and unconfirmed type (type D, leakage of dye from the cut surface, but the source was unclear; n = 13) (Fig. 35.4). In type D pattern, the minor dye leaks could not be visualized by the surgeon and could only be detected by viewing the fluorescent images on the monitor.

A319731_1_En_35_Fig3_HTML.gif


Fig. 35.3
Type C pattern of fluorescence (injured bile duct type). (a) Fluorescent imaging showed two areas of fluorescence corresponding to ducts on the cut surface of the liver (arrows). The lower bile duct was intact (lower arrow). Common bile duct (arrowhead). (b) The lower area of fluorescence corresponded to a partly closed bile duct stump (arrow) under conventional white light illumination. Minor leakage (leaking duct) that is not visible to the surgeons (arrow) is compressed with gauze. (c, d) Fluorescent imaging after the application of gauze shows minor leakage on the cut surface of the liver (arrow), as fluorescence is detected through the gauze (arrowhead). (e) The bile duct stump was repaired by z-suturing with 6–0 nonabsorbable sutures. (f) Fluorescent imaging after the treatment for the minor leakage on the cut surface of the liver (arrow), as fluorescence is not detected through the gauze (arrowhead)


A319731_1_En_35_Fig4_HTML.jpg


Fig. 35.4
Type D pattern of fluorescence (unconfirmed type). (a, b) Fluorescent imaging showed an area of fluorescence on the cut surface of the liver (arrow). (c) Minor leakage of dye is not visible to the surgeons (arrow) under conventional white light illumination. (d) There is major leakage of dye from the cut surface, but the source was unclear

No repair was performed in patients with types A and B patterns of fluorescence. In patients with type C pattern, repair was performed by z-suturing with 6–0 nonabsorbable sutures in 42 patients (1.5 ± 0.8 (mean ± SD) sutures per patient) and by ligation with 4–0 absorbable sutures in the remaining nine patients (1.0 ± 0.2 ligatures per patient). In the three patients with type D pattern, fibrin sealant was applied to the area showing leakage (two pieces of 2.0 × 1.5 cm sealant per patient).


35.3.2 Postoperative Bile Leakage


Postoperative bile leakage occurred in seven of the 132 hepatectomy patients (5 %) and persisted for a median period of 6 weeks (range, 2–15 weeks). Table 35.1 shows the incidence of each pattern of fluorescence. The rate of type D pattern differed from the rates of the other three patterns.


Table 35.1
Incidence of postoperative bile leakage for each pattern of fluorescence











 
Indocyanine green fluorescent imaging

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Nov 20, 2016 | Posted by in GENERAL RADIOLOGY | Comments Off on Intraoperative Indocyanine Green Fluorescent Imaging for Prevention of Bile Leakage After Hepatic Resection

Full access? Get Clinical Tree

Get Clinical Tree app for offline access