Techniques

Nov 15, 2016 by in MUSCULOSKELETAL IMAGING Comments Off on Techniques

Category Intrafractional variations Interfractional variations Systematic Random Systematic Random Variation of CTV size Physiological processes Physiological processes Tumor reduction or swelling Physiological processes Variation of CTV position Change in treatment…

read more

Other Indications

Nov 15, 2016 by in MUSCULOSKELETAL IMAGING Comments Off on Other Indications

Institute/trial [reference no.] No. of patients (risk group) Schedule Free from PSA failure Toxicities Median follow-up SHARP trial [6, 7] 40 (Low) 33.5 Gy/5 fr 93 % at 5 years 1 acute Gr…

read more

International Experience

Nov 15, 2016 by in MUSCULOSKELETAL IMAGING Comments Off on International Experience

Authors (year) Target Patients (Pts) Staging Equipment PTV margin Dose escalation Outcomes MTD RD LF Failure Pattern Whyte RI, et al. (Stanford Univ. & Cleveland clinic) (2003) [3] 1–5 cm in…

read more

Japanese Experiences

Nov 15, 2016 by in MUSCULOSKELETAL IMAGING Comments Off on Japanese Experiences

© Springer Japan 2015Yasushi Nagata (ed.)Stereotactic Body Radiation Therapy10.1007/978-4-431-54883-6_11 11. Japanese Experiences Masaki Kokubo1  , Yasushi Nagata2, Rikiya Onimaru3 and Masahiro Hiraoka4 (1) Department of Radiation Oncology, Kobe City Medical Center General Hospital, 2-1-1, Minatojima-Minamimachi, Chuo, Kobe 650-0047, Japan (2)…

read more

Introduction and History of Stereotactic Body Radiation Therapy (SBRT)

Nov 15, 2016 by in MUSCULOSKELETAL IMAGING Comments Off on Introduction and History of Stereotactic Body Radiation Therapy (SBRT)

© Springer Japan 2015Yasushi Nagata (ed.)Stereotactic Body Radiation Therapy10.1007/978-4-431-54883-6_1 1. Introduction and History of Stereotactic Body Radiation Therapy (SBRT) Yasushi Nagata1   (1) Department of Radiation Oncology, Hiroshima University Hospital, Hiroshima, Japan     Yasushi Nagata Email:…

read more

Quality Assurance (QA)

Nov 15, 2016 by in MUSCULOSKELETAL IMAGING Comments Off on Quality Assurance (QA)

Procedure Tolerance Daily Tests Mechanical Laser localization 1 mm Distance indicator (at isocenter) 2 mm Collimator size indicator : both jaws and MLC 1 mm Safety Stereotactic interlocks (lockout): cone size, backup jaws…

read more